6
c. ProbeUsed
SelectAUTO,Fe,orNoFeunderPROBEUSEDintheOPTIONSMenuusingthearrowbuttonsandthe
SELECTsoftkey.
IntheAUTOmode,themeterautomaticallyactivatestheprobemeasurementmethod(ferrousornon
ferrous)basedonthemetalsubstratethatisbeingmeasured.Whentheprobeis
placedonamagnetic
substrateitwillworkinthemagneticinductionmode.Iftheprobeisplacedonanonferrousmetalit
willworkintheeddycurrentmode.
IntheFerrous(Fe)ModetheMagneticinductionmeasurementmodeisactivated.
IntheNonFerrous(NoFe)Mode
theeddycurrentmeasurementmodeisactivated.
d. Unitsofmeasureselections
Selectmm,µm,ormilsunderUNITSETTINGintheOPTIONSMenuusingthearrowbuttonsandthe
SELECTsoftkey(mm=millimeters;µm=micrometers;mils=mm*2.54/100)
e. Backlight
SelectONorOFFunderBACKLIGHTintheOPTIONSMenuusingthearrowbuttonsandtheSELECTsoft
key.IfOFFisselectedtheLCDbacklightingwillbecompletelydisabled.IfONisselected,theusercan
turnthelightonoroffusingthebacklightbutton(CANCELbutton).
f. LCDStatistics
SelectAVERAGE,MINIMUM,MAXIMUM,orSDEV(StandardDeviation)underLCDSTATISTICSinthe
OPTIONSMenuusingthearrowbuttonsandtheSELECTsoftkey.Thisselectiondetermineswhich
statisticisshownasdefaultontheLCDdisplay.
g. AutoPowerOFF
SelectENABLEorDISABLEunderAUTOPOWEROFFintheOPTIONSMenuusingthearrowbuttonsand
theSELECTsoftkey.Whenenabled,themeterautomaticallyswitchesOFFafter3minutesofinactivity.
Whendisabled,themeterwillonlyswitchOFFwithabuttonpressorwhenthebatterypoweris
weak.
LIMITMenu
HighandLowAlarmLimitscanbesetwhenanAlarmLimitisreachedthemeterdisplaysanAlarmAlerticon
(HforHighAlarmandLforLowAlarm).
1. PresstheMENUbuttontoaccesstheprogrammingmenu
2. UsertheDOWNARROWbuttontoscrolldownto
LIMITandpressSELECT
3. PressSELECTagaintochooseLIMITSETTING
4. PressSELECTagaintochooseHIGHLIMIT
5. UsertheARROWBUTTONStosetaHighAlarmvalue
6. PressOKtostorethelimitandpressBACKtoreturntothemenu
7. Performthesamesteps
fortheLOWLIMIT
8. UsetheDELETELIMITparametertoclearAlarmLimitvalues