User Manual
Oyster-10 pH/mV/Temperature Meter
Introduction
Congratulations on your purchase of Extech's Oyster Series splash-proof meter. This
microprocessor-based device with tactile buttons simultaneously displays pH or mV and
Temperature (
o
C/
o
F selectable). 25 labeled readings can be conveniently stored for later
recall. The Oyster 10 is battery operated, portable, and includes a hinged display that
can be adjusted to any viewing angle. In addition, the neck-strap offers hands-free
operation. Advanced features include diagnostic displays, MIN/MAX readings, audible
alerts, 5-point buffer recognition, & Auto/Manual temperature compensation. This meter
is shipped fully tested and calibrated and, with proper use, will provide years of reliable
service.
Vista de pagina 0
1 2 3 4 5 6 7 8

Indice de contenidos

Pagina 1 - Introduction

User Manual Oyster-10 pH/mV/Temperature Meter Introduction Congratulations on your purchase of Extech's O

Pagina 2 - Oyster-10, V1.8 11/07

Oyster-10, V1.8 11/07 2 Specifications Range (Resolution) Accuracy mV ± 1000 mV DC (1 mV) + 2mV Temperature 32.0 to 212.0oF (0.0 to 99.9oC) +

Pagina 3 - Calibration and Operation

Oyster-10, V1.8 11/07 3 pH Overview pH is a unit of measure (ranging from 0 to 14pH) indicating the degree of acidity or alkalinity of a solution. pH

Pagina 4

Oyster-10, V1.8 11/07 4 pH Measurement and Temperature Compensation 1. Use the pH/mV button to select the pH mode. The unit of measure appears on the

Pagina 5 - MAX/MIN Display

Oyster-10, V1.8 11/07 5 MAX/MIN Display 1. The meter memorizes the highest (MAX) and lowest (MIN) readings when prompted. 2. Press and hold the MAX/

Pagina 6

Oyster-10, V1.8 11/07 6 Maintenance and Troubleshooting Electrode Cleaning The type of electrode coating that occurs over time will determine the clea

Pagina 7

Oyster-10, V1.8 11/07 7 Troubleshooting an Electrode Problem Symptom Cause Recommended Solution Clogged Junction Soak in 4.07 M Potassium Chloride

Pagina 8 - WARRANTY

Oyster-10, V1.8 11/07 8  Support line (781) 890-7440 Technical Support: Extension 200; E-mail: [email protected] Repair & Returns: Extension 21