VB300-en-US_V1.4 8/14
10
BatteryReplacement
The VB300 does not lose its stored readings when the battery expires or when the
battery is replaced. The data logging process will howeverbe stopped and cannot be
re‐started until the battery has been replaced and the logged data has been
downloadedtoPC.
Thebatterylifeis
ratedforapprox.1000hours.Ifindoubt,replacethebatterybefore
logging critical data. Use only 3.6V lithium batteries. Before replacing the battery,
removethedataloggerfromthePC.
Note: When the VB300 is connected to the USB port, its battery discharges at a
higher rate than normal. To
conserve battery life, do not leave the VB300
connectedtotheUSBportforaprolongedperiodoftime.
1. Remove the two Phillips screwsidentified with the two smaller arrows in the
diagrambelow.
2. Slidethebatterycompartmentcoverinthedirectionofthelargerarrow.
3. Replace
thebatterywithoneofthesametype(3.6Vlithium,½AA1.0Ah,Saft
LS‐14250orequiv).
4. Closethebatterycompartmentandtightenwiththetwoscrews.
WARNING: Handle lithium batteries carefully, observe warnings on battery casing.
Disposeofinaccordancewithlocalregulations.