3 407355 Ver. 1.02 9/99
3. METER DESCRIPTION
1. Microphone
2. LCD Display
3. Power ON/OFF key
4. RUN / Left Arrow key
5. SPL, DOSE, TIME Mode and
Left Arrow key
6. Event / Up Arrow key
7. Real time clock / Right Arrow key
8. Reset key (clears Event data)
9. Calibration screw
10. Battery Cover
11. Belt clip
12. PC interface connector
4. DISPLAY DESCRIPTION
1. %DOSE icon
2. Peak detector (> 140dBA)
icon
3. TIME display icon
4. Sound Level measurement
icon
5. High level warning icon (>
115dBA)
6. Event register E1-E5
7. Battery health icon
8. Fast response time icon
9. Slow response time icon
10. Noise accumulation icon
11. Measurement pause (hold) icon
5. DOSE CONSIDERATIONS
DOSE is a parameter used to quantify noise exposure over a period of time. 100% DOSE is
considered to be a 90 dBA noise level over an 8 hour period (in the US, for example). This
value of DOSE is known as the CRITERION. Other Criteria are used, for example 100%
DOSE = 85dBA for some other countries. To see how this works, use the 90dBA example
above and apply it the following.
1. A person experiencing a noise level of 90dBA would receive a 50% DOSE in 4 hours.
2.
If the noise level was 95dBA (5dB higher), a 100% DOSE would be accumulated in only
4 hours. This is because of what’s known as the Exchange Rate whereby a 5dB
increase (selectable) in noise level corresponds to a doubling of noise energy.
3.
The Criterion Level, Exchange Rate and Threshold level are determined by the
applicable national noise standard. Always check the local regulations before using the
meter.