18 MG300-en-US_v2.2 04/15
Meg OHMS
Terminal Voltage Range Resolution Accuracy Test current Short circuit
current
125V (0%~+10%) 0.125~4.000 MΩ 0.001MΩ +(2%+10) 1mA @ load
125kΩ
≤1mA
4.001~40.00 MΩ 0.01MΩ +(2%+10)
40.01~400.0 MΩ 0.1MΩ +(4%+5)
400.1~4000 MΩ 1MΩ +(5%+5)
250V (0%~+10%) 0.250~4.000 MΩ 0.001MΩ +(2%+10) 1mA @ load
250kΩ
≤1mA
4.001~40.00 MΩ 0.01MΩ +(2%+10)
40.01~400.0 MΩ 0.1MΩ +(3%+5)
400.1~4000 MΩ 1MΩ +(4%+5)
500V (0%~+10%) 0.500~4.000 MΩ 0.001MΩ +(2%+10) 1mA @ load
500kΩ
≤1mA
4.001~40.00 MΩ 0.01MΩ +(2%+10)
40.01~400.0 MΩ 0.1MΩ +(2%+5)
400.1~4000 MΩ 1MΩ +(4%+5)
1000V
(0%~+10%)
1.000~4.000 MΩ 0.001MΩ +(3%+10) 1mA @ load
1MΩ
≤1mA
4.001~40.00 MΩ 0.01MΩ +(2%+10)
40.01~400.0 MΩ 0.1MΩ +(2%+5)
400.1~4000 MΩ 1MΩ +(4%+5)
Note: Accuracy specifications consist of two elements:
(% reading) – This is the accuracy of the measurement circuit.
(+ digits) – This is the accuracy of the analog to digital converter.