Extech Instruments Corporation EICMEX505-CP January 2008 Revision: Initial Release JTM-97a CALIBRATION PROCEDURE Revision
Revision #: Initial Page #: EX505 Calibration Procedure 2 of 8 Procedure Number: EICMEX505-CP Table of Contents
Revision #: Initial Page #: EX505 Calibration Procedure 3 of 8 Procedure Number: EICMEX505-CP 1.4. Document Security…………………………………………………...
Revision #: Initial Page #: EX505 Calibration Procedure 4 of 8 Procedure Number: EICMEX505-CP 1.3. Scope / Field of Application This procedure will
Revision #: Initial Page #: EX505 Calibration Procedure 5 of 8 Procedure Number: EICMEX505-CP 3.10. If any of the readings were outside of the spec
Revision #: Initial Page #: EX505 Calibration Procedure 6 of 8 Procedure Number: EICMEX505-CP Appendix: A CHECKING PROCEDURE Page 1
Revision #: Initial Page #: EX505 Calibration Procedure 7 of 8 Procedure Number: EICMEX505-CP 1. VISUAL - Check for cosmetic problems, scratches,
Revision #: Initial Page #: EX505 Calibration Procedure 8 of 8 Procedure Number: EICMEX505-CP Extech Instruments Corporation Capacitance 40.00 nF