Revision #:
Initial
Page #:
EX505 Calibration Procedure
7 of 8
Procedure Number: EICMEX505-CP
1. VISUAL - Check for cosmetic problems, scratches, wrong labels, loose labels, etc.
2. Insert 1 each "9V" battery and observe correct polarity.
3. Power "ON" and check for any missing display segments.
4. Test the following functions:
FUNCTION RANGE INPUT TOLERANCE READING NOTES
DC Volts 400.0 mV 350.0 mV +/- (0.5%rdg+2d) 348.0 – 352.0
DC Volts 4.000 V 3.500 V +/- (1.2%rdg+2d) 3.456 – 3.544
DC Volts 40.00 V 35.00 V +/- (1.2%rdg+2d) 34.56 – 35.44
DC Volts 400.0 V 350.0 V +/- (1.2%rdg+2d) 345.6 – 354.4
DC Volts 1000.0 V 950.0 V +/- (1.5%rdg+2d) 934 – 966
AC Volts 400.0 mV 350.0 mV +/- (2.0%rdg+10d) 342.0 – 358.
0 manually select mV
AC Volts 4.000 V 3.500 V +/- (2.0%rdg+5d) 3.425 – 3.575 auto range
AC Volts 40.00 V 35.00 V +/- (2.0%rdg+5d) 34.25 – 35.75
AC Volts 400.0 V 350.0 V +/- (2.0%rdg+5d) 342.5 – 357.5
AC Volts 1000.0 V 950.0 V +/- (2.5%rdg+5d) 921 – 979
DC Amps 400.0 uA 350.0 uA +/- (1.0%rdg+3d) 346.2 – 353.8
DC Amps 4000 uA 3.500 mA +/- (1.5%rdg+3d) 3.444 – 3.556
DC Amps 40.00 mA 35.00 mA +/- (1.5%rdg+3d) 34.44 – 35.56
DC Amps 400.0 mA 350.0 mA +/- (1.5%rdg+3d) 344.4 – 355.6
DC Amps 10.00 A 7.50 A +/- (2.5%rdg+5d) 7.26 – 7.74
AC Amps 400.0 uA 350.0 uA +/- (2.5%rdg+10d) 340.2 – 359.8
AC Amps 4000 uA 3.500 mA +/- (2.5%rdg+5d) 3407 – 3593
AC Amps 40.00 mA 35.00 mA +/- (2.5%rdg+5d) 34.07 – 35.93
AC Amps 400.0 mA 350.0 mA +/- (2.5%rdg+5d) 340.7 – 359.3
AC Amps 10.00 A 7.50 A +/- (3.0%rdg+7d) 7.20 – 7.80
Resistance 400.0 350.0 +/- (1.2%rdg+4d) 345.4 – 354.6
Resistance 4.000 K 3.500 K +/- (1.0%rdg+2d) 3.463 – 3.537
Resistance 40.00 K 35.00 K +/- (1.2%rdg+2d) 34.56 – 35.44
Resistance 400.0 K 350.0 K +/- (1.2%rdg+2d) 345.6 – 354.4
Resistance 4.000 M 3.500 M +/- (1.2%rdg+2d) 3.456 – 3.544
Resistance 40.00 M 35.00 M +/- (2.0%rdg+3d) 34.27 – 35.73
CHECKING PROCEDURE
Page 2 of 2
Model# EX505
True RMS Industrial Multimeter
FUNCTION RANGE INPUT TOLERANCE READING NOTES
Extech Instruments Corporation