Extech Instruments EX830 Guía de usuario, Página 1: User’s Guide1 Extech Instruments EX830 Guía de usuario, Página 2: User’s Guide2 Extech Instruments EX830 Guía de usuario, Página 3: Introduction3 Extech Instruments EX830 Guía de usuario, Página 4: CAUTIONS4 Extech Instruments EX830 Guía de usuario, Página 5: Description5 Extech Instruments EX830 Guía de usuario, Página 6: Operation6 Extech Instruments EX830 Guía de usuario, Página 7: Resistance Measurements7 Extech Instruments EX830 Guía de usuario, Página 8: Continuity Measurements8 Extech Instruments EX830 Guía de usuario, Página 9: Continuity Measurements9 Extech Instruments EX830 Guía de usuario, Página 10: Units Switch10 Extech Instruments EX830 Guía de usuario, Página 11: Maintenance11 Extech Instruments EX830 Guía de usuario, Página 12: Specifications12 Extech Instruments EX830 Guía de usuario, Página 13: Specifications13 Extech Instruments EX830 Guía de usuario, Página 14: WARRANTY/DISCLAIMER14 Extech Instruments EX830 Guía de usuario, Página 15: Shop online at omega.com15
EX830-EU-EN-V5.2-3/12
8

Non-Contact InfraRed Temperature Measurements

1.
Set the function switch to the
IR Temp
position.
2.
Aim the infrared sensor (rear of the meter) at the surface
to be measured.
3.
Press the
button in the center of the rotary function
switch to turn on the laser pointer and identify the
surface spot to be measured.
4.
The area of the surface to be measured must be larger
than the spot size as determined by the distance to spot
size specification.
5.
Read the temperature in the display.
Note:
See the “Temperature Units” paragraph to select
o
F or
o
C
WARNING: Do not directly view or direct the laser pointer at an eye.
Low power visible lasers do not normally present a hazard, but may
present some potential for hazard if viewed directly for extended periods
of time.
IR Spot to Distance Diagram
The 8:1 spot to distance ratio determines the size of the measured surface area with respect to the
distance the meter is held away from the surface.
IR Measurement Notes
1.
The object under test should be larger than the spot (target) size calculated by the field of view
diagram.
2.
If the surface of the object under test is covered with frost, oil, grime, etc., clean before taking
measurements.
3.
If an object's surface is highly reflective, apply masking tape or flat black paint to the surface
before measuring.
4.
The meter may not make accurate measurements through transparent surfaces such as glass.
5.
Steam, dust, smoke, etc. can obscure measurements.
6.
To find a hot spot, aim the meter outside the area of interest then scan across (in an up and
down motion) until the hot spot is located.
10cm
1.25cm
20cm
2.5cm
60cm
7.5cm

Distance to object

Diameter of spot

Archivo PDF · ·

Imprimir

Imprimir documento

Compartiendo

Preguntas sobre Extech Instruments EX830

Todavía no hay preguntas sobre este manual. Haga la primera.

Hacer una pregunta

Respondemos con el manual e indicamos la página. La pregunta se publica en el sitio después de revisarla.

Todas las preguntas y respuestas ›

Productos y manuales relacionados para Multímetros Extech Instruments EX830