GroundResistance(Autorange)Specifications
Range Resolution Accuracy(ofreading)
0.025to0.250 0.002 ±(1.5%+0.05
0.250to1.000 0.02 ±(1.5%+0.05
1.001to9.999 0.02 ±(1.5%+0.1
10.00to50.00 0.04 ±(2.0%+0.3
50.01to99.99 0.04 ±(2.0%+0.5
100.0to200.0 0.4 ±(3.0%+1.0
200.1to400.0 2 ±(5.0%+5
400.0to600.0 5 ±(10%+10
600.1to1500 20±20%
AccuracyNote:Loopresistancenoninductive,externalfield<50A/m,externalelectricalfield
<1V/m,conductorcentered
Ground/LeakageCurrentSpecifications
Range Accuracy(%ofreading)
0.300to1.000mA ±(2.0%±0.05mA)
1.00to10.00mA ±(2.0%±0.03mA)
10.0to100.0mA ±(2.0%±0.3mA)
100to1000mA ±(2.0%±0.3mA)
Autoranging
0.20to4.00A ±(2.0%±0.03A)
4.00to30.00A ±(3.0%±0.03A)
Manualrange
Notes:50/60Hzbandwidth,TrueRMSsensing,CrestFactor<3.0
HighandLowAlarm(programmable)
Range Resolution
HighAlarm 01510 1
LowAlarm 01510 1